請用此 Handle URI 來引用此文件: http://hdl.handle.net/11455/34918
標題: alpha-Cyclodextrin-modified infrared chemical sensing system that utilizes enzymatic reactions for the determination of glucose
作者: Yang, J.
楊吉斯
Liang, S.C.
關鍵字: infrared sensor
Fr-IR spectroscopy
glucose
enzymatic reaction
GOx
volatile organic-compounds
enhanced raman-scattering
tunable
diode-lasers
chlorinated hydrocarbons
aqueous-solutions
affinity
sensor
spectroscopy
oxidation
water
interference
期刊/報告no:: Analytica Chimica Acta, Volume 537, Issue 1-2, Page(s) 385-392.
摘要: A new infrared (IR) sensing system based on attenuated total reflection (ATR) to determine the levels of glucose in aqueous solutions was developed in this work. To increase the selectivity of the system, we employed glucose oxidase (GOx) to convert glucose into gluconic acid, which was the actual species detected by the IR-ATR sensor. Because gluconic acid is highly polar and was produced generally in low concentrations, the surface of the IR-ATR sensor was modified with cyclodextrin to increase the sensitivity of its detection. As a result, the sensitivity increased roughly by 100 times relative to that of the unmodified ATR. To simplify the detection system, we immobilized GOx onto the surfaces of glass beads and packed them into a column to function as an enzyme reactor. For comparison, GOx was also immobilized onto the internal surface of fused silica tubing. Glucose can be converted to gluconic acid effectively when passed through a flow system and using either packed or open tubular columns. To optimize the performance of the developed system, we investigated the effects of the pH of the sample solutions and the flow rate used to pump glucose through the system. Our results indicate that the optimal pH for the system was ca. 5; a low pH was required for high-efficiency detection of gluconic acid by the CD phase, but it lowered the efficiency of the enzymatic reactions. The flow rate exerted a small influence on the analytical signals over the examined range of flow rates because of the limited rates of the enzymatic reactions. Under the optimal conditions, the linear regression coefficients in the standard curve can be > 0.99 for a concentration range of 1-50 mM and a detection time of 2 min. In addition, our developed system can be regenerated effectively using aqueous solutions at high pH. (c) 2005 Elsevier B.V. All rights reserved.
URI: http://hdl.handle.net/11455/34918
ISSN: 0003-2670
文章連結: http://dx.doi.org/10.1016/j.aca.2005.01.025
顯示於類別:化學系所

文件中的檔案:
沒有與此文件相關的檔案。


在 DSpace 系統中的文件,除了特別指名其著作權條款之外,均受到著作權保護,並且保留所有的權利。