標題: Acute Response in vivo of a Fiber-Optic Sensor for Continuous Glucose Monitoring from Canine Studies on Point Accuracy
作者: Liao, K.C.
張仕杰
Chang, S.C.
Chiu, C.Y.
Chou, Y.H.
廖國智
關鍵字: in vivo glucose monitoring
fiber-optic sensor
point accuracy
Clarke
error grid analysis
calibration
loop insulin delivery
complications
design
期刊/報告no:: Sensors, Volume 10, Issue 8, Page(s) 7789-7802.
摘要: The objective of this study was to evaluate the acute response of Sencil (TM), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measure the interstitial glucose concentration during the ovariohysterectomy procedure. The data was acquired every 10 seconds after initiation, and was compared to the concentration of venous plasma glucose sampled during the surgery procedures for accuracy of agreement analysis. In the four trials with a range of 71-297 mg/dL plasma glucose, the collected 21 pairs of ISF readings from the Sencil (TM) and the plasma reference showed superior dispersion of residue values than the conventional system, and a linear correlation (the Pearson correlation coefficient is 0.9288 and the y-intercept is 14.22 mg/dL). The MAD (17.6 mg/dL) and RMAD (16.16%) of Sencil (TM) measurements were in the comparable range of the conventional system. The Clarke error grid analysis indicated that 100% of the paired points were in the clinically acceptable zone A (61.9%) and B (38.1%).
URI: http://hdl.handle.net/11455/46499
ISSN: 1424-8220
文章連結: http://dx.doi.org/10.3390/s100807789
顯示於類別:生醫工程研究所

文件中的檔案:
沒有與此文件相關的檔案。


在 DSpace 系統中的文件,除了特別指名其著作權條款之外,均受到著作權保護,並且保留所有的權利。