摘要: 
A sensitive voltammetric method is developed for the determination of parathion (phi NO2) using a Nafion(R)-coated glassy carbon electrode. In this method, parathion is first irreversibly reduced from phi NO2 to phi NHOH. The reversible peaks at around +0.33 V (versus Ag/AgCl) corresponding to a two-electron oxidation/reduction of hydroxylamine (phi NHOH) to the nitroso (phi NO) derivative were used for detection with square-wave voltammetry. The experimental parameters, such as, pH, film thickness, preconcentration potential, preconcentration time, and square-wave voltammetric parameters were optimized. Using this method, a linear calibration curve for parathion was obtained up to 15 mu M range in pH 1.1 citrate buffer solution with a detection limit (S/N = 3) of 50 nM. (C) 1999 Elsevier Science B.V. All rights reserved.