Please use this identifier to cite or link to this item: http://hdl.handle.net/11455/70795
標題: Scanning waveform analysis of a room-temperature-probe SQUID NDE system
作者: Chien, J.J.
Lin, I.S.
Horng, H.E.
Hong, C.Y.
Yang, S.Y.
Yang, H.C.
關鍵字: Room-temperature probe;Waveform;Superconducting quantum interference;device (SQUID);t-c squid;quantum interference devices;high-spatial-resolution;nondestructive evaluation;unshielded environment;pickup coil;hts-squid;microscope;samples;flaws
Project: Ndt & E International
期刊/報告no:: Ndt & E International, Volume 43, Issue 7, Page(s) 586-590.
摘要: 
A room-temperature-probe SQUID NDE system with high mobility and excellent spatial resolution has been developed for detecting the internal cracks of metallic material. Here, the sensing probe is composed of quadruple excitation coils and double D-shape pickup coils. It is connected to the input coil surrounding the SQUID sensor to transfer a magnetic field of eddy current. Following the previous demonstration of its good spatial resolution and high sensitivity, a signal mechanism for the scanning waveform is further proposed in this study to explain the scanning non-directionality of the system and analyze the internal cracks. Both fine and wide single cracks and multi-cracks with dense and spare intervals are valid. Moreover, the model is in good agreement with the test results. (C) 2010 Elsevier Ltd. All rights reserved.